The attached are WhiteNoise TFs taken last night and early this morning for the H2 SUS ETMY M0 and R0 masses. For both chains, the Long, Trans, Pitch, & Yaw resonances are vivid above the white noise excitations. The Vertical and Roll DoFs for both masses had very low coherence for various amplitudes. A tuned matlab transfer function measurement is running since early morning for both masses. DTT templates in SVN: ~/SusSVN/sus/trunk/QUAD/H2/ETMY/SAGM0/Data/2012-02-09_M0_*_WhiteNoise*.xml ~/SusSVN/sus/trunk/QUAD/H2/ETMY/SAGR0/Data/2012-02-09_R0_*_WhiteNoise*.xml Plots: ~/SusSVN/sus/trunk/QUAD/H2/ETMY/SAG*/Results/2012-02-09_*_WhiteNoise*.pdf