Summary: I performed 8 coherent H1L1 CBC hardware injections. They were all successful. Waveforms: Waveforms are in the hardware injection SVN: https://daqsvn.ligo-la.caltech.edu/svn/injection/hwinj/Details/Inspiral/${IFO}/coherentbbh*_1126259455_${IFO}.out Waveform parameter files are in the hardware injection SVN: https://daqsvn.ligo-la.caltech.edu/svn/injection/hwinj/Details/Inspiral/coherentbbh*_1126259455.xml Schedule: The following was appended to the schedule: 1128124325 1 1.0 coherentbbh1_1126259455_ 1128125225 1 1.0 coherentbbh2_1126259455_ 1128126125 1 1.0 coherentbbh3_1126259455_ 1128127025 1 1.0 coherentbbh4_1126259455_ 1128127925 1 1.0 coherentbbh5_1126259455_ 1128128825 1 1.0 coherentbbh7_1126259455_ 1128129725 1 1.0 coherentbbh8_1126259455_ 1128130625 1 1.0 coherentbbh9_1126259455_ Recall we had the issue before where injections were being cancelled since they triggered an EM alert for testing. Peter S. documented what needed to be changed in aLog 22163. I reset ${IFO}:INJ-CAL_TINJ_PAUSE and ${IFO}:INJ-CAL_EXTTRIG_ALERT_TIME between injections so that they were not cancelled. It worked. The gracedb entries for these injections are H190051 to H190068. I've attached omegascans of the injections.