duty cycle sept 17th to Oct 2nd, downtime Oct2-3rd

Motivated by our locking difficulties at the end of last week, I downloaded guardian state data for the last two weeks to look into our locklosses durring the CARM offset reduction.

To start with the good news, the script that I use to do this automatically makes a few figures that tell us about our duty cycle, and we are doing well overall. As usual, I am only looking at the guardian state, and not considering any intent bits.  My reaasoning is that the hardest part is getting and keeping the IFO locked, and that is currently the important limit to our duty cycle. 

The main weakness in our lock acquistion sequence now seems to be the early steps of CARM offset reduction, as we saw yesterday. 

Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 22:12, Tuesday 06 October 2015 (22294)

As requested, I've made a few plots which bettter show how individual locks contribute to our total low noise time.  The first attached plot is a repeat of the histogram of lock times from above, with the lower panel sshowing the sum of the low noise time accumlated in all the locks in that bin of the histogram.  The second attached plot shows the accumlated low noise time, theere is a point for each lock stretch with the length of the lock on the horizontal axis and the cumulative low noise time on the vertical axis.  

The message of both plots is that we accumulate a lot of our time durring a small number of long locks.  The secondd plot shows that in the 3 longest locks (out of 33 total) we accumlated 133 hours of low noise time out of 312 total hours.

Images attached to this comment
sheila.dwyer@LIGO.ORG - 15:24, Wednesday 07 October 2015 (22310)

I also had a look at relocking times for this period based on a request from Keita. 

The minimum relocking time was 20 minutes (this is measured from the time of lockloss to the time that the guardian arrives in nominal low noise), the median was 53 and the maximum was 6 hours and 20 minutes (long stretch of downtime on the evening of Oct 2nd is not included here).

A histogram is attached.

Images attached to this comment