H1-L1 coherence over O1 so far
Here are the observed H1-L1 coherences as of October 12, 2015, as calculated by stochmon:
https://ldas-jobs.ligo.caltech.edu/~thomas.callister/stochmonO1/stochmon.html

Some plots attached.

100 mHz resolution. 1E-5 is the expected level of coherence due to uncorrelated noise.
Frequency, coherence
33.6 Hz, 1E-3
34.6 Hz, 2E-2
35.4 Hz, 1E-3
42.5 Hz, 6E-4 (barely above background, could be a noise fluctuation)
64.0 Hz, 2E-4 (barely above background, could be a noise fluctuation)
74.0 Hz, 5E-4 (barely above background, could be a noise fluctuation)
128.0 Hz, 1E-2
256.0 Hz, 2E-3
331.8 Hz, 3E-4 (barely above background, could be a noise fluctuation)
351.1 Hz, 4E-4 (barely above background, could be a noise fluctuation)
559.9 Hz, 2E-4 (barely above background, could be a noise fluctuation)
736.0 Hz, 3E-3 (23rd harmonic of 32 Hz)
1904.0 Hz, 2E-3 (119th harmonic of 16 Hz)
2160.0 Hz, 4E-4 (135th harmonnic of 16 Hz)
2376.0 Hz, 4E-3 (297th harmonic of 8 Hz)
2768.0 Hz, 7E-3 (173rd harmonic of 16 Hz)
2984.0 Hz, 7E-4 (373rd harmonic of 8 Hz)
3001.2 Hz, 1E-1 (calibration lines)
4928.0 Hz, 3E-3 (77th harmonic of 64 Hz)
5016.0 Hz, 3E-3 (627th harmonic of 8 Hz)
5144.0 Hz, 2E-1 (largest coherence, 643rd harmonic of 8 Hz)
5616.0 Hz, 2E-3 (351st harmonic of 16 Hz)
5624.1 Hz, 6E-3 


0.25 Hz, 4E-6 is the expected level of coherence due to uncorrelated noise.
Frequency, coherence
34.25 to 35.25 Hz, 1E-3
36 to 37 Hz, 1.5E-3
37.25 Hz, 2E-3
42.5 Hz, 2E-4
128.0 Hz, 3E-3
256 Hz, 6E-4
1904.0 Hz, 4E-4 (119th harmonic of 16 Hz)
3552.25 Hz, 2E-4
4163.75 Hz, 5E-3 Hz
4576.0 Hz, 2E-4 (143rd harmonic of 32 Hz)
5016.0 Hz, 7E-4 (627th harmonic of 8 Hz)
5144.0 Hz, 1E-1 (largest coherence, 643rd harmonic of 8 Hz)
5584.5 Hz, 2E-2
5655.5 Hz, 1E-3
Images attached to this report
Comments related to this report
nelson.christensen@LIGO.ORG - 20:03, Monday 12 October 2015 (22452)DetChar
Continues the study on the H1-L1 h(t) coherence to date for O1.

0.001 Hz resolution, 2E-3 is the expected level of coherence due to uncorrelated noise.

0.5 Hz harmonics are seen in the H1-L1 coherence, averaged over all of O1 to data. See figures. Various other lines to be seen as well.

Frequency, Coherence
18.50 Hz, 3E-2
19.00 Hz, 3E-2
19.50 Hz, 4E-2
20.00 Hz, 3.5E-2
23.50 Hz, 2.5E-2
25.00 Hz, 3E-2
25.50 Hz, 2.5E-2
32.00 Hz, 2E-2
36.7 Hz, 3E-2
42.5 Hz, 3E-2
43.0 Hz, 5E-2
48.0 Hz, 4E-2
61.243 Hz, 2E-2
71.0 Hz, 7E-2
71.50 Hz, 3E-2
100.00 Hz, 7E-2
128.00 Hz, 5E-1
256.00 Hz, 2E-1
896.0 Hz, 3.5E-2
952.0 Hz, 7E-2
1080.0 Hz, 1.5E-1
Images attached to this comment