Displaying report 1-1 of 1.
Reports until 11:11, Tuesday 20 October 2015
H1 General (TCS)
nutsinee.kijbunchoo@LIGO.ORG - posted 11:11, Tuesday 20 October 2015 (22675)
CO2 laser recovered after power outage

RF Oscillator is set at 40.68 MHz.

DC power supplies set at 29.99V and 28A. Output reads 30.00V and 0.75A for TCSX, 30.00V and 0.78A for TCSY.

The rotation stage has been reset for both TCSX and Y (searched for home, then go-to power). TCSX output is now 0.229W.

TCSY laser is currently turned off. It's been off for over a month. I checked with Evan and he said it should be left off for now since no one will play with it anytime soon. To turn it on just hit "ON" button on the CO2Y medm screen. The controller box is ready to go.