MIDY vacuum leak

Kyle will report more but here is a plot of pressure for the last 4 hours. We are not sure how permanent this "fix" might be but the magnitude of recovery suggests that this is the dominant or only leak in this volume.

Images attached to this report
Comments related to this report
john.worden@LIGO.ORG - 10:00, Thursday 22 October 2015 (22745)

This morning the pressure is at 1.6e-9 torr.  We had been at 4.6e-9 torr. At these pressures the main ion pump may have 1000 l/s pumping speed (guess) for air. Given this the air leak was on the order of 3e-6 tl/s.

The largest signal seen on the helium leak detector was ~ 1e-7 tl/s for helium. However, the leak detector was only sampling a fraction of the helium as we are open to the beam tube on both sides.