Kyle, Gerardo We retested the feed-through flanges which had previously passed leak testing but which had since had their cable "back-shells" replaced ((12) cables on (6) 4.5" flanges). We did this to verify that the act of disconnecting and reconnecting the cables required to replace the back-shells had not resulted in any "new" leaks -> We found no new leaks. Indicated helium baseline remained ~6 x 10-10 torr*L/sec throughout testing. Next, we retested the 4.5" CF blank which had initially been found to leak at 1.1 x 10-7 torr*L/sec earlier but which had been tightened and found to be fixed (see earlier entries) -> No leaks Finally, we retested the known leaker (dual triax) and verified it to be leaking at 2 x 10-7 torr*L/sec (note: slight difference between current value and value previously reported is due to calibrating using a different source of calibration gas, external vs. internal cal-gas source). We plan on venting the YBM next week and to replace the leaking triax flange as well as reflanging the feedthrough suspected of having "bottomed-out" flange screws (see earlier entries).