The full report is posted on the O1 DQ Shift wiki
-
The overall duty cycle was 39.1%, this was mostly due to violin modes and microseisms on Monday, then maintenance on Tuesday, but we finished strong with 79% duty cycle on Wednesday!
-
The small amount of time that was locked on Monday was clean.
-
The Tuesday locks had a few loud glitches that had high enough to X^2-weighted SNR to be troublesome to the BBH search but we tracked that down to DAC overflow glitches.
-
After looking at the OmegaScans, it looked like the glitches will be picked up in the autogating portion of the offline searches.
-
Wednesday was a good day and there were no loud glitches that aregoing to mess up BNS or BBH searches. There was a dip in the range starting around 11:30 that was correlated with increased seismic noise seen in the BLRMS. HVETO also concentrated most of its efforts in the later part of the day.