There was some suspicion that the whitening gain for some of the ACB PDs became bad, but I have confirmed on the floor that the gain setting on the MEDM screen is reflected correctly in the diode amplifier box (D1301017) over the entire gain range for all baffle PDs for IX, IY, EX and EY.
I've followed the test procedure E1400003.