J. Kissel, K. Izumi, N. Kijbunchoo
We've gathered new DARMOGLTF measurements to continue our long-term investigating on the slow evolution of the calbriation parameters. Based on what I see in DTT alone, the uncorrected loop gain and pcal to CAL-CS transfer function remains within 5% and 5 [deg] of the reference model. Very good! I attach screen shots of the raw measurements and a conlog of the relevant settings, but more detailed analysis to come.
The new DTT results live here:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O1/H1/Measurements/DARMOLGTFs/2015-10-28_H1_DARM_OLGTF_7to1200Hz.xml
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O1/H1/Measurements/PCAL/2015-10-28_PCALY2DARMTF_7to1200Hz.xml
and have been exported to the following:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O1/H1/Measurements/DARMOLGTFs
1 / (1 + G) = 2015-10-28_H1_DARM_OLGTF_7to1200Hz_A_ETMYL3LOCKIN2_B_ETMYL3LOCKEXC_coh.txt
2015-10-28_H1_DARM_OLGTF_7to1200Hz_A_ETMYL3LOCKIN2_B_ETMYL3LOCKEXC_tf.txt
-G = 2015-10-28_H1_DARM_OLGTF_7to1200Hz_A_ETMYL3LOCKIN2_B_ETMYL3LOCKIN1_coh.txt
2015-10-28_H1_DARM_OLGTF_7to1200Hz_A_ETMYL3LOCKIN2_B_ETMYL3LOCKIN1_tf.txt
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O1/H1/Measurements/PCAL/
C / (1 + G) = 2015-10-28_PCALY2DARMTF_7to1200Hz_A_PCALRX_B_DARMIN1_coh.txt
2015-10-28_PCALY2DARMTF_7to1200Hz_A_PCALRX_B_DARMIN1_tf.txt
Sudarshan, Darkhan, RickS
Using data from the SLM tool for 67 minutes just before the measurement of the Pcal2Darm TF, we calculated the TF for the 1 kHz and 3 kHz Pcal lines.
The plots attached below show:
1) The "kappas"
2) The measured TF, with coherence and the fractional uncertainty
3) The data for the 1 kHz line and comparison with TF values interpolated from the measured TF data
4) The data for the 3 kHz line
At 1 kHz, the transfer coefficient calculated from the SLM data differes from the interpolated coefficient from the meassured transfer function by 1.01% in amplutude and 0.15 deg. in phase.
The script and plots will be submitted to the SVN tomorrow and I will add a comment giving their location when they are.