Displaying report 1-1 of 1.
Reports until 12:01, Thursday 12 November 2015
H1 INJ
sudarshan.karki@LIGO.ORG - posted 12:01, Thursday 12 November 2015 (23343)
Inverse anti-imaging filter on pcal hardware injection path

SudarshanK, ChrisB, DaveB

We installed inverse anti-imaging filter on pcal hardware injection path, channel H1CAL-PINJX_HARDWARE, at FM3. The SDF screen and ODC bit are changed accordingly.

This filter will produce a better estimate of the  hardware injection  at higher frequencies.