Reports until 12:36, Tuesday 17 November 2015
H1 AOS
jason.oberling@LIGO.ORG - posted 12:36, Tuesday 17 November 2015 - last comment - 07:28, Wednesday 18 November 2015(23473)
H1 Optical Lever Re-zeroed

Today I re-zeroed all of the active H1 optical levers.  The table below shows their pitch/yaw values before re-zeroing and the values as they read when I finished each oplev (there could be relaxation in the motors causing a slow, slight drift, especially in the BS and HAM2 oplevs which are zeroed with piezo motors steering a 2" turning mirror).  This closes work permit #5606.

Optical Lever Old (µrad) New (µrad)
Pitch Yaw Pitch Yaw
ETMx -2.4 -2.2 -0.2 0.3
ETMy -1.3 -6.7 0.1 0.0
ITMx -10.2 -8.2 -0.2 0.0
ITMy 7.9 -2.6 -0.2 -0.0
PR3 -1.1 -2.9 0.0 0.0
SR3 10.1 -5.7 0.0 0.2
BS -13.3 -9.2 -0.4 0.4
HAM2 -20.6 -39.0 0.1 -0.1
Comments related to this report
jeffrey.kissel@LIGO.ORG - 07:28, Wednesday 18 November 2015 (23501)DetChar, ISC, SUS
Tagging SUS, DetChar, and ISC for future reference.