Today I re-zeroed all of the active H1 optical levers. The table below shows their pitch/yaw values before re-zeroing and the values as they read when I finished each oplev (there could be relaxation in the motors causing a slow, slight drift, especially in the BS and HAM2 oplevs which are zeroed with piezo motors steering a 2" turning mirror). This closes work permit #5606.
Optical Lever | Old (µrad) | New (µrad) | ||
Pitch | Yaw | Pitch | Yaw | |
ETMx | -2.4 | -2.2 | -0.2 | 0.3 |
ETMy | -1.3 | -6.7 | 0.1 | 0.0 |
ITMx | -10.2 | -8.2 | -0.2 | 0.0 |
ITMy | 7.9 | -2.6 | -0.2 | -0.0 |
PR3 | -1.1 | -2.9 | 0.0 | 0.0 |
SR3 | 10.1 | -5.7 | 0.0 | 0.2 |
BS | -13.3 | -9.2 | -0.4 | 0.4 |
HAM2 | -20.6 | -39.0 | 0.1 | -0.1 |