Displaying report 1-1 of 1.
Reports until 12:19, Wednesday 18 November 2015
H1 AOS
patrick.thomas@LIGO.ORG - posted 12:19, Wednesday 18 November 2015 (23514)
Cleared IMC WFS histories
Cheryl, Patrick

It appears that earthquake drove the IMC WFS to a position where the IMC could no longer lock. I cleared the histories on all degrees of freedom on the IMC WFS master screen using the small CH button in the bottom right of each filter bank. This brought MC2 back to a position where the IMC could lock. Plots attached.
Non-image files attached to this report
Displaying report 1-1 of 1.