We would like to know the range of counts we have available for doing hardware injections with PCAL.
Hardware injections are filtered with an inverse actuation filter to get from strain to counts, this is the CAL-PINJX_HARDWARE filterbank.
Here I've filtered waveforms with the CAL-PINJX_HARDWARE filterbank and see how many counts they require. I've attached plots of the time series in counts.
Last Thursday a filter was added to compensate for the upsampling that is done in the analog electronics. This seems to have a significant effect on the counts. I did a test with the coherentbbh0 waveform (used in hardware injections tests before) and saw that it increased its counts by a factor of ~3. I've attached plots for before and after adding this filter.
Here I report the max number of counts from the waveforms:
* CW injections: ~1500 counts
* 1.4-1.4 non-spinning SEOBNRv2 waveform (target SNR is ~15): 7e7 counts
* 25-25 non-spinning SEOBNRv2 waveform (target SNR is ~15): ~10000 counts
* coherentbbh0 waveform from hardware injections tests (recovered SNR is ~18-23): ~20000 counts
Note these SNRs are approximates and can change depending on sky location. It was quoted that perhaps the CW injections could change by a factor of 2.
There is also a higher-frequency line that gets injected in PCALX and we need to take this into consideration as well.