Reports until 11:16, Tuesday 24 November 2015
H1 INJ (CAL, DetChar, INJ)
christopher.biwer@LIGO.ORG - posted 11:16, Tuesday 24 November 2015 (23701)
Turn off 3kHz PCALX line and switch CW injections to use PCALX
Chris B., Keith R.

Summary:

Chris turned off the 3kHz PCALX line.

Keith updated the CW injections to use PCALX and Chris tested it. Everything went well.

Turned off the 2kHz line:

The 3kHz PCALX line turned off at 18:37 UTC. I accepted the change for H1:CAL-PCALX_OSC_SUM_ON that changes the value from 1 to 0 on the SDF.

Updated psinject config file:

Keith R. updated psinject to inject into CAL-PINJX_CW_EXC. To do this Keith updated psinject_config.tcl and changed the line " set channel CAL-INJ_CW_EXC" to "set channel CAL-PINJX_CW_EXC". Chris did the following commands to update:
cd /home/hinj/Details
svn up psinject_config.tcl

Tested CW injections on PCALX:

CW injections were off at LHO already and the monit web interface has been disabled. See aLog 23666.

So I turned off the CAL-PINJX_CW filterbank and did a test of psinject. To do the test I did:
cd /home/hinj/Details/bin
./start_psinject
./check_psinject
./stop_psinject

I checked that the CW injections were going to the correct channel. Everything looked fine.

I then turned off psinject. I turned the CAL-PINJX_CW filterbank on.