keita.kawabe@LIGO.ORG - posted 09:21, Wednesday 07 March 2012 (2374)
TMS picomotor cable was swapped
On Tuesday 06/Mar/2012, we pulled one of D1000223 cables (S1104079, used for picomotors, connecting the feedthrough and the connector bracket on the ISI table) and replaced with  S1104077, which is from a different bake batch. Now all picomotors work.
We tested the bad cable using a cable tester. It tested fine at first, but when I wiggled the cable near the connector I was able to repeatedly make it fail.
Richard wants to test this cable using HV mode of the tester.
 
History of this signal chain (also read https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=2350):