Yesterday, we measured the resistances of the 12 actuators on ISI-BSC8. The table below shows the measured resistance when the ISI was:
- in the staging building
- on the teststand in the LVEA
- in the chamber (March 6, 2012)
Field cables in the LVEA are about 1 ohm. Resistances measured after the important drive are similar to those measured when the ISI was on the LVEA teststand. There are no obvious signs of damage on these actuators.
Staging building | LVEA Teststand | Chamber after large drive | |
Actuators | Resistance (ohm) | Resistance (ohm) | Resistance (ohm) |
ST1 H1 | 6.4 | 7.5 | 7.4 |
ST1 H2 | 6.6 | 7.4 | 7.4 |
ST1 H3 | 6.6 | Not measured | 7.4 |
ST1 V1 | 6.4 | 7.5 | 7.5 |
ST1 V2 | 6.6 | 7.4 | 7.5 |
ST1 V3 | 6.6 | Not measured | 7.5 |
ST2 H1 | 10.3 | 11.3 | 11.3 |
ST2 H2 | 10.3 | 11.1 | 11.1 |
ST2 H3 | 10.1 | Not measured | 11.1 |
ST2 V1 | 10.1 | 11.1 | 11.2 |
ST2 V2 | 10.3 | 11.1 | 11.2 |
ST2 V3 | 10.5 | Not measured | 11.3 |