two plots attached: Dec. 1st and Dec. 5th - all while H1 is locked in Low Noise
Dec. 1st plot shows BS optical lever sum glitching by 200+ counts.
Dec. 5th plot of the same signals shows ETMY optical lever sum glitching 50-70 counts, and ETMX optical lever sum gltiching 200+ counts, but BS sum is now quiet and +/-25 counts from the mean sum value.
Is this glitching, or an artifact of something else, and could this be effecting the IFO's locking?
Red herring... Variation in signal is a small percentage of the sum.