J. Kissel Another collection of our bi-monthly(-ish) assessment of the calibration and DARMOLGTF. Details will be processed later including analysis of time-dependence. The front-end, CAL-CS, DELTAL_EXTERNAL calibration (and therefore the low-latency GDS pipeline output) is still within the desired precision. I attach screenshots of the measurements for posterity. Note, this time, I've updated the reference to be the most recent sweep, taken Dec 01 2015, (see LHO aLOG 23883, also yet to be processed.) DARM OLGTF Template: /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O1/H1/Measurements/DARMOLGTFs/2015-12-21_H1_DARM_OLGTF_7to1200Hz.xml Corresponding Exports: /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O1/H1/Measurements/DARMOLGTFs/ 2015-12-21_H1_DARM_OLGTF_7to1200Hz_A_ETMYL3LOCKIN2_B_ETMYL3LOCKEXC_coh.txt 2015-12-21_H1_DARM_OLGTF_7to1200Hz_A_ETMYL3LOCKIN2_B_ETMYL3LOCKEXC_tf.txt 2015-12-21_H1_DARM_OLGTF_7to1200Hz_A_ETMYL3LOCKIN2_B_ETMYL3LOCKIN1_coh.txt 2015-12-21_H1_DARM_OLGTF_7to1200Hz_A_ETMYL3LOCKIN2_B_ETMYL3LOCKIN1_tf.txt PCAL2DARM TFs: /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O1/H1/Measurements/PCAL/2015-12-21_PCALY2DARMTF_7to1200Hz.xml Coresponding Exports: 2015-12-21_PCALY2DARMTF_7to1200Hz_A_PCALRX_B_DARMIN1_coh.txt 2015-12-21_PCALY2DARMTF_7to1200Hz_A_PCALRX_B_DARMIN1_tf.txt