H1 SEI (OpsInfo)
jim.warner@LIGO.ORG - posted 00:03, Saturday 26 December 2015 (24483)
Briefly out of Observe, EX ISI was ringing up, blends switched

About 2 hours ago the EX ISI started ringing up again. As the situation was getting worse over the last few minutes, I decided it was time to switch blends. Attached plot shows the StripTool I was watching with the EX CPS locations and IMC-F (the gray trace). I switched the Y DOF blend first (green), it changed smoothly, but didn't reduce the motion. Then I changed the X blend (blue), which settled things down pretty much immediately. I thought it was interesting that IMC-F seemed to lead the swings of the ISI, ie IMC-F seemed to change direction a little before the ISI did. Don't know what that means, but interesting.

I add for the operators: If an ISI is ringing up, and you change the beam direction blend to the 90 blends (X for EX, Y for EY), you should change the perpendicular blend as well. ASC probably doesn't care, ALS doesn't, but 20 micron motion of the tables is bad just on principle.

Images attached to this report