H1 General (SEI)
travis.sadecki@LIGO.ORG - posted 20:20, Saturday 02 January 2016 - last comment - 20:25, Saturday 02 January 2016(24645)
Temporary drop in range related to tidal limits

We noticed that the range had dropped and the ASC loops had rung up.  Guardian DIAG was also notifying us that the tidal control was nearing its limits at both ETMx and ETMy.  While digging through the aLog for solutions (and finding only speculative fixes involving switching of ISI blends), things seemed to have calmed down and returned to normal.  However, plotting the TIDAL_CTRL signals for both X and Y shows that the control signal is much larger than we have seen in quite some time.  The first plot is a 2 day plot of the TIDAL_CTRL signal for both arms, and the second is a 30 day plot.

Images attached to this report
Comments related to this report
travis.sadecki@LIGO.ORG - 20:25, Saturday 02 January 2016 (24646)

We are not sure we understand why Guardian DIAG is throwing the 'within 10% of its limit' notification.  The control signal for both arms is ~300 while the limit for both arms is set to 700. 