Displaying report 1-1 of 1.
Reports until 13:37, Tuesday 05 January 2016
H1 SEI
hugh.radkins@LIGO.ORG - posted 13:37, Tuesday 05 January 2016 (24710)
LHO STS2s Mass Positions Centered-Finally

Attached is a plot of the STS2's Mass Positions.  Finally got them all centered (something below 5000cts)

Found that ETMY did not have the Binary I/O cable from the Bin Output to the STS2 Interface so this was not going to re-center remotely anyway.  Vincent pulled that cable but we don't quite understand.  I can tell you though that the recentering does not work via the medm/cable but the signal select and period select toggling does work...  We've left that cable unplugged and did the centering at ETMY via the front panel button.

Images attached to this report
Displaying report 1-1 of 1.