Ops Eve Summary: 00:00-08:00UTC, 16:00-23:59PT
State of H1: unlocked
Commissioners: Jenne, Sheila (fellow), JeffK
Incoming Operator: JimW
Shift Summary:
-
H1 unclocked for calibration measurements at EY when I arrived
-
I tried locking and ended up doing an initial alignment
-
One H1 lock about half way through the shift with good range to start, but low POP_A_LF
-
Sheila and Jenne worked of increasing POP_A_LF using PRM, which did improve the signal
-
Lockloss, not alignment work, based on the ASC signals before lock loss
-
PRM was moved, however it's new position was reset during locking (old p/y offsets -1412.7/515.3, new p/y offsets -1405.3/530.1)
-
Relocking revealed some Guardian issues with PRMI to DRMI transition
-
Currently H1 has lcoked and made it through DRMI but did not make it to Engage ASC part 1
-
POPAIR RF18 and RF90 are popping to 100 or more, but at this time, no locked DRMI
-
PRM is in it's old alignment position at 8:12UTC
-
Handing off to JimW and Sheila is here