Reports until 16:46, Friday 15 January 2016
LHO General
thomas.shaffer@LIGO.ORG - posted 16:46, Friday 15 January 2016 - last comment - 17:34, Friday 15 January 2016(24978)
Lockloss 00:25 UTC

Robert was running inj at the time, but I'm not sure if that was it or not. Still investigating.

Comments related to this report
thomas.shaffer@LIGO.ORG - 17:34, Friday 15 January 2016 (24980)

Attaching the generic lockloss plot. Nothing evident here.

ISI BS St1 and St2 WD tripped during the lockloss, so I have also attached the ACT plot. Robert was inj on ISI BS at the time so I suspect it was this inj that caused the lockloss.

Images attached to this comment