Displaying report 1-1 of 1.
Reports until 20:10, Friday 15 January 2016
H1 ISC (OpsInfo)
evan.hall@LIGO.ORG - posted 20:10, Friday 15 January 2016 (24983)
IMC-F offloading limits increased

Sheila, Jenne, Evan

We had another instance of the integrated ALS tidal offsets causing the IMC-F offloading to bump up against the rail.

We weren't sure how to turn on the bleed feature of the ALS filters, so we instead doubled the digital limiter on the COMM control filters.

Displaying report 1-1 of 1.