Displaying report 1-1 of 1.
Reports until 10:03, Thursday 28 January 2016
LHO VE
kyle.ryan@LIGO.ORG - posted 10:03, Thursday 28 January 2016 - last comment - 11:02, Friday 29 January 2016(25221)
X2-8 BT ion pump shut off ("shut down" on display) @ 0630 hrs. local
0915 -0955 hrs. local -> To and from X-end 

Pump won't start on either HV channel, gets to 500V then limits at 500ma -> Tempted to increase the default arc limit of 10 per minute but will consult with GammaVacuum first
Comments related to this report
kyle.ryan@LIGO.ORG - 10:55, Thursday 28 January 2016 (25225)
1020 - 1040 hrs local -> To and from X-end 

Consulted with Gamma Vacuum -> They advised not exceeding 20 arc/min setting -> I tried 15 - no help -> I also unplugged the HV cable from the controller but left the SafeConn connector connected and then enabled the HV which then came up to 7500V without a problem - meaning the controller is likely OK - more to come
scott.mccormick@LIGO.ORG - 11:02, Friday 29 January 2016 (25239)
What type of HV cable is in service? Gamma's cable or RG58 RED? Did you pull both types of cables to have one as a back up? 
Displaying report 1-1 of 1.