Alastair, Evan
We drove a line at 24.1 Hz in the TCS X AOM to see if it would show up in DARM.
A 0.003 ct excitation into TCS-ITMX_CO2_AOM_OUT_GAIN is enough to create a line in DARM with an SNR of ~5 for a 10-second FFT.
2015-01-30 22:52:15 to 22:57:15 Z.