Displaying report 1-1 of 1.
Reports until 15:02, Wednesday 03 February 2016
H1 ISC (ISC)
jenne.driggers@LIGO.ORG - posted 15:02, Wednesday 03 February 2016 - last comment - 04:27, Sunday 07 February 2016(25364)
5dB attenuators on new 90MHz distribution box outputs

[Jenne, Cao]

What was once the 90 MHz local oscillator for just the ASAIR 90 WFS now goes to a distribution amplifier box.  3 of the outputs of that board now go to the local oscillator inputs for each of ASAIR_90, AS_A_90 and AS_B_90. The local oscillator inputs want 10dBm each, so we measured the output of the distribution box, and each output was 15dBm.  So, we put 5dB RF attenuators on each spigot.  Next up, we'll lock DRMI and look at phasing.

Comments related to this report
daniel.sigg@LIGO.ORG - 04:27, Sunday 07 February 2016 (25432)

Looks like the RF amp gets overdriven at the input. The outputs should be around 13dBm.

Displaying report 1-1 of 1.