Reports until 10:33, Tuesday 09 February 2016
H1 CDS
patrick.thomas@LIGO.ORG - posted 10:33, Tuesday 09 February 2016 - last comment - 11:21, Tuesday 09 February 2016(25459)
Added IscRfCAsAmp90m channels to h1ecatc1
WP 5727

Yesterday Daniel and I changed h1ecatc1 PLC2 to replace IscRfCSpareAmp[1] of IscRfCSpareAmp[1..5] with IscRfCAsAmp90m. This morning I rescanned PLC2 in the system manager and added the links for IscRfCAsAmp90mIn and the lost links for the remaining IscRfCSpareAmpIn (now IscRfCSpareAmpIn[1..4]).

IscRfCAsAmp90mIn.OutputMon linked to Corner Chassis 6 L1 Channel 4
IscRfCAsAmp90mIn.PowerOk linked to Corner Chassis 6 L3 Channel 4

IscRfCSpareAmpIn[1..4].OutputMon linked to Corner Chassis 6 L2 Channel 1..4
IscRfCSpareAmpIn[1..4].PowerOk linked to Corner Chassis 6 L4 Channel 1..4

The wiring diagram also needs to be updated to show that IscRfCTcsAom40mIn.OutputMon is linked to Corner Chassis 6 L1 Channel 3 and IscRfCTcsAom40mIn.PowerOk is linked to Corner Chassis 6 L3 Channel 3.

Since everything was restarted I burtrestored PLC1, PLC2 and PLC3 to 6:10 this morning local time.
Comments related to this report
patrick.thomas@LIGO.ORG - 11:21, Tuesday 09 February 2016 (25463)
I updated the channel link list E1201049-v9.