Reports until 16:44, Wednesday 17 February 2016
H1 SUS
filiberto.clara@LIGO.ORG - posted 16:44, Wednesday 17 February 2016 (25599)
TMSX SAT AMP Units Replaced (EX).
This afternoon, we replaced the Sat Amp units for TMSX. New units have the same modifications as the SR3/OMC units. Two decoupling capacitors were placed on the input of the negative regulator (U503), as implemented at LLO. 

C602 0.1uf
C601 10uf

New Unit S1100098 Old Unit S1100168
New Unit S1000292 Old Unit S1100066