Reports until 13:53, Monday 09 April 2012
H2 PSL
michael.rodruck@LIGO.ORG - posted 13:53, Monday 09 April 2012 (2568)
PSL down

Rick, Michael R

We have shut down the PSL after discovering the diode chiller flow had dropped from 22 lpm to 13 lpm over the weekend. This appears to be caused by a collapsed water filter (photo is attached). The current filters in use are 1 micron, but we plan on swapping them out with 5 micron filters this week.

Images attached to this report