Jenne, Sheila, Hang
We wrote a piece of code to automate the measurement of ASC sensing matrix (using method described in alog 25648). It is available at
/opt/rtcds/userapps/release/isc/h1/scripts/SenMtrx
This is a complementary code to the existing one. The new one does not connect to the nds server but only uses ezca, so it may be more stable. On the other hand it can only do elementary measurement right now.
=============================================================================================================
We used this code to measure the sensing matrix with only DRMI locked. The values are
| Optic | WFS | ct/ct | W/rad |
| PR3 | REFL_9A_I | -3.2e-1 | -6.1e+1 |
| PR3 | REFL_45A_I | 1.2e-1 | 2.3e+1 |
| PR3 | REFL_9B_I | -2.2e-1 | -4.1e+1 |
| PR3 | REFL_45B_I | 9.5e-2 | 1.8e+1 |
| PR2 | REFL_9A_I | -1.1e0 | -2.1e+2 |
| PR2 | REFL_45A_I | 4.8e-1 | 9.1e+1 |
| PR2 | REFL_9B_I | -7.8e-1 | -1.5e+2 |
| PR2 | REFL_45B_I | 3.4e-1 | 6.5e+1 |
| IM4 | REFL_9A_I | -1.5e-1 | -2.8e+1 |
| IM4 | REFL_45A_I | 4.8e-2 | 9.1e0 |
| IM4 | REFL_9B_I | 9.5e-2 | 1.8e+1 |
| IM4 | REFL_45B_I | -4.6e-2 | -8.6e0 |
| SR3 | AS_36A_I | -1.7e-2 | -1.6e0 |
| SR3 | AS_36A_I | -1.1e-2 | -1.1e0 |
Why are the signals from PR3 so weak and weaker than those from PR2 ? Maybe typos ?
Typo fixed