PSL Weekly 10 day trends and Status
Laser Status:
SysStat is good
Front End power is 32.01W (should be around 30 W)
Frontend Watch is GREEN
HPO Watch is RED

PMC:
It has been locked 12.0 days, 22.0 hr 9.0 minutes (should be days/weeks)
Reflected power is 3.078Watts and PowerSum = 25.24Watts.

FSS:
It has been locked for 0.0 days 0.0 h and 15.0 min (should be days/weeks)
TPD[V] = 1.466V (min 0.9V)

ISS:
The diffracted power is around 8.465% (should be 5-9%)
Last saturation event was 0.0 days 0.0 hours and 15.0 minutes ago (should be days/weeks)

For further in-depth analysis of the trends, please refer to Jason O., Pete K. or Rick S.