Reports until 08:57, Wednesday 16 March 2016
H1 General (PSL)
edmond.merilh@LIGO.ORG - posted 08:57, Wednesday 16 March 2016 (26096)
PSL Weekly status and trends
Laser Status:
SysStat is good
Front End power is 31.72W (should be around 30 W)
Frontend Watch is GREEN
HPO Watch is RED

PMC:
It has been locked 1.0 days, 19.0 hr 34.0 minutes (should be days/weeks)
Reflected power is 3.035Watts and PowerSum = 24.92Watts.

FSS:
It has been locked for 1.0 days 0.0 h and 4.0 min (should be days/weeks)
TPD[V] = 1.501V (min 0.9V)

ISS:
The diffracted power is around 9.648% (should be 5-9%)
Last saturation event was 0.0 days 16.0 hours and 33.0 minutes ago (should be days/weeks)

Please refer further, in depth, analysis of PSL trends to Jason, Peter or Rick.
Images attached to this report