HWSX camera images still glitching after H1HWSMSR reboot

Dave B rebooted the H1HWSMSR machine yesterday. Unfortunately, an investigation of the HWSX images shows that they are still glitching for an unknown reason. 

Images attached to this report
Comments related to this report
kiwamu.izumi@LIGO.ORG - 15:57, Friday 25 March 2016 (26256)

Nutsinee, Kiwamu,

As requested by Aidan, we have restarted the cameras and the computer yesterday after the LHO ISC meeting. We remotely shut the cameras by pressing the buttons on the medm screens. Then we went to the MSR and pressed the reset button to restart the computer. Aidan reported that the corrupted image issue has gone before we shut off the cameras. After rebooting the computer and cameras, we restarted the HWS codes using the old tmux sessions.

Also Aidan discovered that when the images were corrupted it reported an anomalously high pixel mean peak values and in fact it exceeded the theoretical limit with the 12 bits data = 4096 counts. This motivated us to set up a diagnosis test in the DIAG guardian as reported in 26243.