Reports until 14:51, Thursday 24 March 2016
H1 TCS (TCS)
aidan.brooks@LIGO.ORG - posted 14:51, Thursday 24 March 2016 (26237)
HWSX glitch investigation

I investigated the HWS glitching issue in more detail. I ran the following tests to try to reproduce the glitching seen previously:

  1. Run each camera individually with the take command (number of ring buffers = 4): no glitching
  2. Run each camera individually with the take command (number of ring buffers = 1): no glitching
  3. Run both cameras at the same time with the take command (number of ring buffers = 1): no glitching
  4. Run both cameras at the same time with the take command (number of ring buffers = 4): no glitching
  5. Run both Run_HWS_ITMX and Run_HWS_ITMY at the same time: no glitching

Since we couldn't reproduce the glitching after everything restarted, Kiwamu and I decided that we create some Guardian diagnostics to run continuously on the HWS data. The following channels will be monitored and will issue warning in the case of aberrant behaviour: