J. Kissel As per this April HAM6 vent's plan, E1600092, I've checked that the SUS have not been affected by this week's vent activities by measuring all DOFs of all SUS in the chamber -- the OMC and OMs 1 through 3. Attached are the results of the transfer functions compared against the last in-vacuum results. All SUS are A-OK; SUS is a go for chamber close out. New data sets live here: /ligo/svncommon/SusSVN/sus/trunk/OMCS/H1/OMC/SAGM1/Data/2016-04-08*.xml /ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM1/SAGM1/Data/2016-04-08*.xml /ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM2/SAGM1/Data/2016-04-08*.xml /ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM3/SAGM1/Data/2016-04-08*.xml Note, I've improved the OM templates by tuning the drive and adding a boost to the low-frequency drive. The OMC's template didn't need it.