IOT2L IMC REFL unlocked O2 power budget expanded

I've expanded the IOT2L power budget from the one I posted in alog 26390 to include all components in the beam path and maximum power levels (saturation/damage) for critical components.

I've attached summary of the IMC REFL unlocked power budget that includes only the O1 (22W), proposed O2 (50W), and future (200W) power levels.

Two components are highlighted in yellow in the proposed O2 columns:

The full documentation is uploaded to T1600107-v1.