H1 PSL (PSL)
peter.king@LIGO.ORG - posted 05:03, Wednesday 20 April 2016 (26675)
PSL work update
The replacement turbine flow sensor seems to be working okay and without any leaks.

Whilst the laser was down, the dark voltage of the pre-modecleaner locking photodiode
and frequency stabilisation reflection photodiode was measured.
  PMC locking photodiode : 23.5 - 23.7 mV
       FSS RF photodiode : -4.6 - -4.7 mV

    Whilst engaging the PMC, the output of the locking photodiode would saturate and latch
at random intervals, causing the PMC to oscillate.  This photodiode was replaced with a
spare.
  old : S1107847
  new : S1107848
The measured maximum power that could fall incident on the locking photodiode was ~63 mW.
These occurred at ~83 degrees and 265 degrees.  The minimum power was ~1 mW, and this
occurred at 167 degrees and 347 degrees.

    After locking the FSS, ~10.1 mW was measured after the Faraday isolator located behind
the reference cavity.  This was increased to ~30 mW by rotating WP4 from 98 degrees to 86
degrees.

    Had problems with the stability of the power stabilisation.  The free-running power
fluctuations transmitted by the PMC were perhaps at most 3%.  Yet when the power stabilisation
servo was engaged, it was working far harder than this figure would suggest.  Not obvious why.
Perhaps it's a scattered light issue?





Jason, Peter