LHO VE
chandra.romel@LIGO.ORG - posted 17:52, Wednesday 20 April 2016 - last comment - 02:43, Thursday 21 April 2016(26701)
CP6 below 100% full
After 15 hrs, CP6 finally registers a real value on the liquid level read back. Currently at 98.5%. I opened the Dewar valve and set the LLCV to 55% (normally runs between 60-65%). I'll be here for a while working on CAD drawings so I'll monitor it.
Comments related to this report
chandra.romel@LIGO.ORG - 18:14, Wednesday 20 April 2016 (26702)
Magnehelic reads a hair over 40 in. H20 (maybe 40.4?) when liquid level reads back 97.8%. Ratios match what we expect!
chandra.romel@LIGO.ORG - 18:16, Wednesday 20 April 2016 (26704)
Raised LLCV to 57%.
chandra.romel@LIGO.ORG - 21:32, Wednesday 20 April 2016 (26707)
Raised level to 60% valve open (liquid level reads 85.7%).
Images attached to this comment
chandra.romel@LIGO.ORG - 23:42, Wednesday 20 April 2016 (26710)
I was going to manually fill CP6 a bit before I left, but the bypass valve is frozen shut. Liquid level is 80.5% in the red, but LLCV is set to its normal value of 60%. Should be good and ready for some Beckhoff PID action in the morning.
kyle.ryan@LIGO.ORG - 02:43, Thursday 21 April 2016 (26711)
75% @ 0240 hrs. local