Which PSL monitor PDs correspond to which signal, and which DCC document.

Since there seem to be some confusions about which PD has what kind of analog filtering and sent to which channel, here it is.

I'm quite certain about HPO output before AOM (which is "Power monitor PD" in D1002929) and ISS 1st loop monitors, but not that sure about "monitor PD"s in D1002164. E-travellers are incomplete (they don't say which one is installed where), so I'm just listing the nominal values for these "monitor PD"s.

What PD name on D0902114 Circuit type Analog out Transimpedance (Ohm) Analog filter Channel Note Can you find Filter MEDM from sitemap as of this writing?
HPO output before AOM PD1

Power monitor PD, D1002929

DC 3.3k DC  

H1:PSL-PWR_HPL_DC, DC_LF

DC_LF is digital downstream of DC, EPICS output is visible as "Power Monitor PD" on PSL_LASER MEDM.

No
AC 16.5k AC 5k HPF nominal H1:PSL-PWR_HPL_AC   No
ISS 1st loop diodes after PMC ISS_PDA, ISS_PDB

Inner loop diodes, D1001998

"Filt" on the board, "AC" on the box 660 DC

z=[0.0723;2700;0.0707] Hz,

p=[3.3607;130;3.12;2300]

H1:PSL-ISS_PDA and PDB

No dewhite, output is calibrated in volts.

No

H1:PSL-ISS_PDA_DC and PDB_DC

Digitally low-passed version of PDA and PDB, but has a DC gain of 5 so the DC agrees with the analog of "DC" output on the PD box. Yes, from ISS
H1:PSL-ISS_PDA_AC and PDB_AC Dewhitened and AC-coupled version of PDA and PDB, has a gain to match PDA_DC and PDB_DC Yes, from ISS
H1:PSL-ISS_PDA_REL AC-coupled RIN made by PDA_AC/PDA_DC. Yes, from ISS
DC on the box 3.3k DC   N/A   N/A
Frontend output before HPO but after FI PD_AMP

PSL monitor PD, D1002164, T100047

DC on the box 20k DC nominal   H1:PSL-OSC_PD_AMP_DC EPICS output visible as "FRONTEND POWER" on PSL_LASER MEDM. No
AC on the box 100k AC nominal 5k HPF nominal AMP_AC   No
Back-propagation rejected by FI between frontend and HPO PD_ISO PSL monitor PD, D1002164, T100047 DC 750 DC nominal   H1:PSL-OSC_PD_ISO_DC EPICS output visible as "PDISO" on PSL_LASER MEDM. No
AC 3.75k nominal 5k HPF nominal ISO_AC   No
Back-propagating HPO mode leaking from HPO cavity? PD_INT PSL monitor PD,  D1002164, T100047 DC 1k?   H1:PSL-OSC_PD_INT_DC EPICS output visible as "PDINT" on PSL_LASER MEDM. No
AC 5k? 5k HPF nominal INT_AC   No
HPO Brewster plate rejection PD_BP PSL monitor PD, D1002164, T100047 DC 1.5k nominal   H1:PSL-OSC_PD_BP_DC EPICS output visible as "PDBP" on PSL_LASER MEDM. No
AC 7.5k nominal 5k HPF nominal BP_AC   No
Comments related to this report
keita.kawabe@LIGO.ORG - 10:07, Wednesday 11 May 2016 (27113)

I cannot edit the above entry any more, so here is an additional table showing digital filters.

what analog channel model digital
HPO output before AOM DC H1:PSL-PWR_HPL_DC h1pslpmc None
  H1:PSL-PWR_HPL_DC_LP p=0.05
AC H1:PSL-PWR_HPL_AC None
ISS 1st loop diodes after PMC "filt" or AC
(DC-coupled)
H1:PSL-ISS_PDA (and PDB) h1psliss cts/volt conversion factor
  H1:PSL-ISS_PDA_CALI_AC

z= [0.0707, 0.0723], p= [0.3, 0.3] and 2nd order 0.3Hz Butterworth HP in addition to dewhite, DC gain of 5.

  H1:PSL-ISS_PDA_CALI_DC Some random LPF (p=[0.034141, 0.037449, 10430]), DC gain of 5.
    H1:PSL-ISS_PDA_REL None
Frontend output before HPO but after FI DC H1:PSL-OSC_PD_AMP_DC h1pslpmc Gain of 0.00349223
AC H1:PSL-OSC_PD_AMP_AC Gain of 0.000613
Back-propagation rejected by FI between frontend and HPO DC H1:PSL-OSC_PD_ISO_DC Gain set to 1 on May 11 2016 (alog 27112)
AC H1:PSL-OSC_PD_ISO_AC Gain of 1
Back-propagating HPO mode leaking from HPO cavity? DC H1:PSL-OSC_PD_INT_DC Gain set to 1 on May 11 2016 (alog 27112)
AC H1:PSL-OSC_PD_INT_AC Gain of 1
HPO Brewster plate rejection DC H1:PSL-OSC_PD_BP_DC Gain set to 1 on May 11 2016 (alog 27112)
AC H1:PSL-OSC_PD_BP_AC Gain of 1
PMC TRANS ? H1:PSL-PWR_PMC_TRANS Gain of 0.0103, p=0.15
PMC REFL DC ? H1:PSL-PWR_PMC_REFL Gain of -0.0248015, p=0.15
keita.kawabe@LIGO.ORG - 15:44, Tuesday 30 August 2016 (29392)

ISS inner loop (or 1st loop) diode has different filter than written above, it turns out. But 130Hz was a zero, not pole. 2.7k was a pole, not zero.

effective trans impedance = 660 Ohm (that's 0.2*3.3k).

z=[0.0707; 0.0723; 130]

p=[3.12; 3.36; 2.34k; 2.70k]

jeffrey.kissel@LIGO.ORG - 08:45, Wednesday 31 August 2016 (29409)DetChar, IOO, ISC
Tagging DetChar, IOO, and ISC for future reference.
keita.kawabe@LIGO.ORG - 12:52, Thursday 01 September 2016 (29441)

Seems like I was really, really tired, here's a correction of correction. Really sorry for the confusion.

My original table was correct.

Inner loop PD is equivalen of 660Ohm, zp=([0.0707;0.0723;2.7k],[3.12;3.36;130;2.34k]).