ITM L2 noise - 1.5nV/rtHz at 25Hz

Superseeds alogs 26910 and 26924

Bad news: There is lots of MHzish pick-up on the cables to the ITM L2 coils: ~50mVpk
Good news: The ITM L2 coil noise at low freuency is very good: 1.5nV/rtHz at 25Hz, and we might not care about all that pickup.

Details:
The 10Hz harmonics reported in alog 26910 was a measurement problem, generaterd in Rai's preamplifier box (D060205). The cables pick up on the order of 50mVpk at around 1MHz, which was amplified by 100x, causing slew-rate down-conversion.
This was fixed (in the measurment setup) with a 270nF capacitor in prarallel to the 23.8Ohm cable and coil resistance, resulting in a 24.7kHz pole to cut off the cable pick-up.
 
Plot 1 and 2:
ITMX and ITMY coil noise.
Configuration:
- Everything (coil driver, cable, coil) was connected. The breakout box was inserted between coil driver and cable to the satellite amplifier.
- The L2 cois drivers were both is state 3: Acq Off, LP On, which is the run mode. They are never switched for the ITMs.
- The coil driver inputs were left connected to the DAC/AI. I also sent a 100ctpk, 3kHz signal into H1:SUS-ITM[XY]_L2_DRIVEALIGN_L2L_EXC, corresponing to a 2.6ctpk signal on the DAC. I did this to make sure the DAC is as least flipping bits, which raises its noise level.
- A 270nF capacitor was put in parallel to the coil using a pomona box to avoid saturating the D060205-preamp.
- The preamp has a gain of 100. After the preamp a 100Hz low pass was used (1595Ohm and 1uF) to allow the SR785 to run in the lowest noise mode.

Plot 3:
Noise projection assuming incoherent noise and assuming the ETMs behave the same.


Plot 4:
High frequency noise pick-up on the coil cable (coil driver disconnected).
The dominant noise is at ~1MHz, broadband.
Shown are two traces: one in nominal configuratiuon (green), and one with an additional choke on the cable to the coils.

Plot 5:
Scope trace of the high-frequency signal (only the x100 amplifier is used). The signal is made up of ~10msec bursts every ~100msec.

Plot 6:
All 4 coils (without amplifyer) directly connected to the scope. Note that the grounded inputs of the scope slightly change the signal.

Plot 7:
Scope trace with only an antenna connected to the scope. The signal pickup was largest between the racks - I could not trace it to a source yet.

Images attached to this report
Comments related to this report
stefan.ballmer@LIGO.ORG - 21:01, Monday 02 May 2016 (26951)

For reference, the data and matlab code is available at ~controls/sballmer/20160429/plotIt.m

stefan.ballmer@LIGO.ORG - 21:08, Monday 02 May 2016 (26952)

Also, since we probably have to do similar noise checks when we have the IFO back, here is the equipment I used: Top: Ring antenna Bottom, from left to right: choke, breakout card, 270nF parallel capacitor box, Rai's preamp box, 100Hz LP filter, AC coupling for looking at RF on the spectrum analyzer.

Images attached to this comment
richard.mccarthy@LIGO.ORG - 08:28, Tuesday 03 May 2016 (26956)
Plot 4.  With the cable disconnected from the Coil drive the Shield is no longer terminated.  This may contribute to the pickup.
stefan.ballmer@LIGO.ORG - 19:41, Tuesday 03 May 2016 (26986)

The high frequency noise coupling in in plot 4 is mostly common, and shows up because Rai's preamp has no differential sensing.

In the attached plot the noise seen on ITMX coil 1 is plotted, once sensed with Rai's single-ended preamp, once with an SR560 in differential mode.

 

Conclusion: This noise does not show up on the coil current.

However: The same common HF noise pickup seems to be present on all cables. This now makes me worry about the ESD: I suspect the ESD has much less common rejection, because the +400V and -400V comes in on different cables. Moreover, a broadband noise at 1MHz on the ESD will produce noise near DC due to the quadratic nature of the ESD coupling.

Images attached to this comment
stefan.ballmer@LIGO.ORG - 19:45, Tuesday 03 May 2016 (26987)

I also tried to use the antenna to locate the source. The field is strongest in a circle arouind the rack, suggesting the source might not be in the electorics room, but rather just brought in by all the cables.