~1200 hrs. local -> bake of Vertex RGA started
Kyle 

Multiple hiccups delayed the actual start of the bake and, as such, this exercise will now drag into Friday -> I am utilizing a second isolation valve via adding, in series, the turbo+gauge portion of a donor pump cart.  I modified the wiring so that this redundant isolation valve closes on a fore line pressure set point. The nominal fore line isolation valve closes only upon the loss of AC to the scroll pump motor.  