unexplained vacuum gauge signal stepping
After Tuesday activities, noticed a step change in pirani gauge signals. Attached is plot of PT100 (on HAM 1) and PT140 (on BSC4). Note:  PT100 CC gauge keeps turning on and off due to pirani set point and stepping.

PT120 also shows a step in signal. Yesterday Richard & Gerardo pulled the leads on PT120 while measuring the voltage. Voltage read higher valve when lifted from rack.

PT140 pirani trend over 15 days shows a major step from 4/26 (when Beckhoff was installed) to 5/3 when power was switched and system rebooted.
Images attached to this report
Comments related to this report
michael.zucker@LIGO.ORG - 14:24, Thursday 05 May 2016 (27032)

Sounds like the input impedance of the Beckhof ADC is lower than the original VME card ADC. These form a voltage divider with the gage head's output impedance.  I/O impedances  should be listed in the respective instrument datasheets.  There may be a jumper/selection available.

Beyond recalibrating epics records,  good to confirm that the gage heads are happy driving whatever impedance they see full-scale.  

 

kyle.ryan@LIGO.ORG - 19:59, Thursday 05 May 2016 (27040)
Good idea Chandra.  I hadn't even considered looking at the piranis!