kyle.ryan@LIGO.ORG - posted 10:13, Friday 06 May 2016 - last comment - 10:18, Friday 06 May 2016(27043)
0930 hrs. local -> Heating portion of Vertex RGA bakeout complete -> Pumps running until further notice
Notes to my future self -

Soaked @ 120C - 150C for 60 hours.  

Static power to maintain zones at temperature = 405 watts 

Zone specifics 

Turbo + 1.5" valve -> VAR = 38%, 0.18 amp and 200 ohm 
1.5" valve to reducing Tee -> VAR = 44%, 0.48 amp and 136 ohm
2.5" to 1.5" reducer -> VAR = 44%, 0.9 amp and 70 ohm 
2.5" to 1.5" reducing Tee -> VAR = 44%, 1.27 amp and 48 ohm 
Two 1.5" Tee assembly -> VAR = 38%, 0.39 amp and 140 ohm
RGA #1 -> VAR = 44%, 0.88 amp and 70.5 ohm 
RGA #2 + elbow -> VAR = 44%, 0.9 amp and 70 ohm 
2.5" valve -> VAR = 44%, 0.92 amp and 70 ohm 
N2 leak valve -> VAR = 38%, 0.40 amp and 141 ohm 
Kr leak valve -> VAR = 38 %, 0.40 amp and 137 ohm 
Comments related to this report
kyle.ryan@LIGO.ORG - 10:18, Friday 06 May 2016 (27044)
Here "VAR" used as abbreviation for VARIAC and not "VOLT-AMP REACTANCE"