Displaying report 1-1 of 1.
Reports until 17:07, Monday 09 May 2016
H1 CDS
sheila.dwyer@LIGO.ORG - posted 17:07, Monday 09 May 2016 (27077)
medm updates after directory cleaning up durring RCG update

Durring the RCG upgrade last week, the automatically generated medm screens were cleaned up.  Apparently there were some places (ALS end station overview is one example) where we were pointing to screens in the wrong directory since the ISCEND and ALS models were split.  This worked until the directories were cleaned up last week.  

I've added macros for alseamodel, alsebmodel, alexmodel, and alseymodel to the isc macros file (userapps/isc/common/medm/isc_macro.txt, iscex_macro.txt, iscey_macro.txt), and edited ALS_ENDSTATION_OVERVIEW.adl to use this.  There might be other places we need to change the model we are pointing to.

All of this is in svn. 