[Sheila, Jenne, Cheryl]
We found that many optics' drivealign sliders were in the wrong place after last week's RCG upgrade - these are things that we specifically keep out of the SDF system, since they are often changing by small amounts. But, this means that they aren't kept up to date in the safe and down SDF files. Anyhow, they have been restored, and we were able to lock and align the green arms, and were also able to lock the Xarm in IR for the input align step of initial alignment.
Cheryl will attempt the rest of the alignment sequence, so we have a nice IFO for tomorrow.