11:45 am local 1/2 turn open LLCV bypass valve. Took 12:00 min to overfill CP3 today at 20% LLCV open.
Took longer than usual to overfill at 20% open (from nominal 18%) because yesterday during maintenance the CP3 controls were disabled which caused the valve to want to close, except the shims (set to around 18%) kept it open.