I have updated the dtt calibration template for the DARM spectrum that are shown in the control room. This is for correcting for the new IOP decimation filters. I already applied this update to the latest DARM template, H1_DARM_FOM_20160229.xml.
The generation script is checked in the SVN at
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/PreER9/H1/Scripts/ControlRoomCalib/H1DARM_FOM_correction.m
This script returns an ASCII file where the calibration transfer function is written. The ASCII file can be found at
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/PreER9/H1/Scripts/ControlRoomCalib/DARM_FOM_calibration_new_20160512.dat